Spirolab Spirometer
Technical Specifications

Power supply:
rechargeable battery, standard AC
Temperature sensor:
semiconductor 0-45ºC
Flow and volume transducer:
digital turbine
Flow/volume range:
0-16 Ls-1/10L, BTPS
Volume accuracy:
± 3% or 50mL whichever is greater
Flow accuracy:
± 5% or 200mL/s whichever is greater
Dynamic flow resistance:
<0.8 cm H20/L/s@14L/s
Graphic display:
FSTN 240x320
Printer paper:
thermal, 112mm width
Mouthpieces:
30mm external diameter
Serial output:
RS 232, optoisolated
Dimensions:
310 x 205 x 65mm
Weight:
approx. 4 lbs.
Parameters:
FVC, FEV1, FEV1%, FEV6, FEV1/FEV6%, PEF, FEF25-75%, FEF25%, FEF50%, FEF75%, FET, Vext, FVC, FEV1, PEF, PIF, FIVC, FIV1, FIV1%, VC, IVC, ERV, IC, VT, VE, BF, Ti, Te, Ti/Ttot, VT/Ti, MVV

Spirolab meets ATS and ERS standards.

Optional carrying case.